Effect of BiV Pacing on VT and VF

So finally, what is the effect of BiV pacing on VT and VF?

Most electrophysiologists can remember a case or two in which biventricular pacing significantly reduced the amount of ventricular tachycardia. However, while several studies suggested an antiarrhythmic effect, other reports questioned if the dispersion of repolarization and the prolongation of the QT interval associated with BiV pacing may actually have proarrhythmic effects. A recent analysis from the MADIT-CRT investigators[1] shed new light on this clinical conundrum.

In the MADIT-CRT study, 1,820 patients with ischemic or non-ischemic cardiomyopathy (EF <30%), NYHA class I–II and a QRS duration of >130ms were randomized to either BiV-ICD or ICD implantation with a 3:2 randomization. Based on the reduction of the left ventricular end-systolic volume (LVESV) at 1 year, the patients were divided into high responders with >25% reduction (n=529), low responders with <25% reduction in LVESV, apart from the “regular” ICD patients (n=623).

During the two-year follow-up, 13% of patients experienced ventricular arrhythmia after a median time of 1.24 years. Using a categorical analysis, ventricular arrhythmias were most likely in low responders (28%), less frequent in ICD-only patients (21%), and least likely in high responders to BiV pacing (12%) with a p-value of <0.001. In a continuous analysis, each incremental 10% reduction of LVESV reduced the risk of subsequent ventricular arrhythmia by about 20%.

First, the good news: 71% were high responders with a significant reduction of the LVESV. It appears that there is a close correlation between the mechanical and the electrical improvement and that the reduction in HF seems to go hand-in-hand with a decrease in ventricular arrhythmia. However, these positive effects were only seen in patients with positive response to BiV pacing. Therefore, a good response to resynchronization therapy seems to be critical. No information was available in the study about lead placement, but it is reasonable to conclude that every effort should be made in achieving the best possible CS lead location.

The million dollar question is, of course, if those results also apply to patients with class III and IV heart failure symptoms. Unfortunately, this study cannot answer the question, but this question will have to be answered in upcoming trials. For now, one may have to assume that the same observations hold true for more advancing stages of HF.

Interestingly, there was a trend to more ventricular arrhythmias in the low responders versus the ICD patients, raising the question if resynchronization therapy might indeed be proarrhythmic in this subgroup. MADIT-CRT has identified several predictors of high clinical response such as female gender, left bundle branch block, a wider QRS and non-ischemic CMP. Clearly, we should use those clinical criteria — and others that will come forward — for careful patient selection to minimize the group of low responders. My personal feeling is that those will also be eventually reflected in the BiV guidelines to allow an even more tailored approach to BiV implantation.

Reference
1. Barsheshet A, Wang PJ, Moss AJ, et al. Reverse remodeling and the risk of ventricular tachyarrhythmias in the MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial-Cardiac Resynchronization Therapy). J Am Coll Cardiol 2011; 57:2416-2423.

Timm Dickfeld, MD, PhD is the Chief of Electrophysiology at the VA Baltimore, Associate Professor of Medicine at the University of Maryland, and Founder of the Maryland Arrhythmia and Cardiology Imaging Group (MACIG).

Opinions expressed by authors, contributors, and advertisers are their own and not necessarily those of HMP Communications, the editorial staff, or any member of the editorial advisory board. HMP Communications is not responsible for accuracy of dosages given in articles printed herein. The appearance of advertisements in this journal is not a warranty, endorsement or approval of the products or services advertised or of their effectiveness, quality or safety. HMP Communications disclaims responsibility for any injury to persons or property resulting from any ideas or products referred to in the articles or advertisements.

image description image description
Opinions expressed by authors, contributors, and advertisers are their own and not necessarily those of HMP Communications, the editorial staff, or any member of the editorial advisory board. HMP Communications is not responsible for accuracy of dosages given in articles printed herein. The appearance of advertisements in this journal is not a warranty, endorsement or approval of the products or services advertised or of their effectiveness, quality or safety. HMP Communications disclaims responsibility for any injury to persons or property resulting from any ideas or products referred to in the articles or advertisements.


Parker Knightsays: June 23.2011 at 18:00 pm

Does anybody know the effect of the type of AV optimization used on non-responders? Don't think the study required certain optimization methods but hopefully the data was recorded for analysis. Could effect non-responders and frequency of arrhythmias.

Reply to this comment »

Post new comment

  • Web page addresses and e-mail addresses turn into links automatically.
  • Lines and paragraphs break automatically.

More information about formatting options

10 + 7 =
Solve this simple math problem and enter the result. E.g. for 1+3, enter 4.