AcuNav Reprocessing: An Evaluation of Safety, Clinical Effectiveness and Cost Savings
- 2 (Feb. 2006)
- Posted on: 5/1/08
- 0 Comments
- 7480 reads
Image Quality
Since image quality and resolution are perceived to be the most critical clinical aspects of ultrasound, each catheter was subjected to element-by-element testing. To clearly detect element failure or potential degradation after reprocessing, each sample was tested on a commercially available transducer testing platform, FirstCall 2000. This system is a high-speed, portable probe testing system and has a custom fixture for signal acquisition. This equipment was used to measure basic acoustic performance parameters within a population of reprocessed AcuNav catheters. The repeatability of the FirstCall testing was analyzed in a separate study, and was found to be acceptable for all the testing parameters. The following data was collected and analyzed: element sensitivity, capacitance, pulse width, center frequency, fractional bandwidth, weak or dead elements. However, only element sensitivity and capacitance is being reported in this article to limit the scope.
Dielectric Strength
Reprocessed AcuNav catheters were subjected to 100% hipot (high potential) and leakage current testing. In hipot testing, often called the voltage breakdown or dielectric strength test, the AcuNav's outer shaft was electrically stressed at a worse case to withstand electric stress gradients at elevated voltages for prolonged durations. Fundamentally, the test is designed to evaluate how well the outer shaft contains and insulates from voltage leaks and how resistant the insulation is to voltage arcing. As shown in Figure 2, sparking or arcing is characterized by rapid variations in voltage or current that typically results in failure. If kinks, cuts or abrasions are present within the material, an escape path is created for the voltage. Maintaining an adequate dielectric barrier between the electrical power and the patient is critical in terms of power surges. In case of malfunctioning equipment or unsteady electrical current, breached dielectric insulation can result in patient or end user shock or electrocution. The testing sequence includes submerging the catheter shaft into a saline bath enclosed in a custom-designed test vessel. Elevated voltages and currents are transmitted through the length of the catheter from the handle to the distal transducer. Dielectric testing measurements were performed in accordance to the industry-accepted standard UL 2601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety.7 All tested catheters were found to be within the specification set by the industry-acceptable standard.
Leakage Current
1. Hynes BJ, Mart C, Artman S, et al. Role of intracardiac ultrasound in interventional electrophysiology. Curr Op Cardiol 2004;19:52-57.
2. Weigang B, Moore W, Gessert J, et al. The Methods and Effects of Transducer Degradation on Image Quality and the Clinical Efficacy of Diagnostic Sonography. J Diagn Med Sonography 2003;19:3-13.
3. Eisner L, Brown MR, Modi D. Leakage Current Standards Simplified. MD&DI July 2004.
4. QuadTech. Electrical Safety Testing Reference Guide. 1998.5. Shafer M. Sonic Tech, Inc. Analysis Report on Alliance Medical Reprocessing of AcuNav Transducers: Analysis of FirstCallâ„¢ 2000 repeatability with AcuNav transducers, 2004.
6. IEC 60601-1, -Medical Electrical Equipment - Part 1: General Requirements for Safety and Essential Performance- (Geneva: International Electrotechnical Commission, 1995).
7. UL 2601-1, -Medical Electrical Equipment, Part 1: General Requirements for Safety- (Northbrook, IL: Underwriters Laboratories, 2003).
8. ANSI/AAMI ES1:1993, -Safe Current Limits for Electromedical Apparatus- (Arlington, VA: AAMI, 1993).
9. Report on acoustical measurements conducted for Alliance Medical, Inc., Sonora Medical Systems, Report number 561, pg14.
10. AAMI TIR 30, A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices, 2004.
11. ANSI/AAMI ST35-1991. Good Hospital Practice: Handling and Biological Decontamination of Reusable Medical Devices. ANSI/AAMI.
12. Food and Drug Administration. 1996. Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance, Office of Device Evaluation.
13. Avitall B, Khan M, Krum D, et al. Repeated use of ablation catheters: A prospective study. J Am Coll Cardiol 1993;22:1367-1372.












Post new comment