Prolonged temporary pacing is associated with frequent complications. We describe a patient with aortic endocarditis and acquired tri-fascicular block in whom back-up pacing was indicated. Using a Seldinger technique via a subclavian approach, a permanent active-fixation lead was positioned in the right ventricle. The lead was tunnelled subcutaneously for 6 cm, and the proximal end was connected to a standard single-chamber pulse generator. The procedure was well tolerated, and over a period of four months there were no complications or infection. The PR interval subsequently reduced in duration to 200 ms and as no episodes of AV block had occurred, the lead was easily removed with retraction of the helix and gentle traction.


       Temporary pacemakers are associated with high complication rates, and if pacing is required for more than 24 hours, the need for repeat intervention to reposition the lead or re-site the venous access is relatively commonplace.1–4 We describe a case where an alternative approach was taken to anticipated prolonged temporary pacing.

Figure 1.
Image of the right pectoral region showing the proximal puncture site (P) where the lead has been secured to the subcutaneous tissues with the pacing lead cuff (not seen). The exit point of the lead (E) is around 6 cm caudal to the proximal puncture site. Suture (S) from the recently removed tunnelled central venous catheter are seen medial to the pacing lead exit point.


Case Report
       A 79-year-old man presented to hospital with a history of weight loss and fever. He had undergone aortic valve replacement with a porcine prosthesis two years prior for infective endocarditis. On examination he appeared anemic, cachectic and was febrile. There was a soft murmur in the aortic area. ECG showed left bundle branch block (LBBB), left axis deviation and borderline first degree atrioventricular block (PR 200 ms). Transthoracic and trans-esophageal echocardiography showed evidence of an aortic root abscess and prosthetic aortic valve endocarditis. Methicillin resistant Staphylococcus Aureus (MRSA) was subsequently isolated in blood cultures, and appropriate antibiotic therapy was commenced.
       On serial ECGs over a period of 10 days, the PR interval gradually increased to 280 ms. In view of this and the pre-existing left bundle branch block, prophylactic temporary pacing was felt appropriate due to the theoretical risk of the development of high-grade AV block. As a tunnelled central venous line had already been placed via the right subclavian vein, access was attempted via the left internal jugular vein. This provided adequate pacing thresholds, but the temporary pacing lead repeatedly dislodged requiring re-manipulation, and a stable position could not be achieved.
       As a prolonged period of prophylactic pacing was anticipated but implantation of a permanent system was not possible due to sepsis, we elected to implant a hybrid pacing system. Under local anaesthesia the right subclavian vein was directly punctured, and using the Seldinger technique, a standard 7 French introducer sheath was used to pass an active fixation permanent pacemaker lead (Medtronic CapSureFix Novus) to the right ventricle. The acute pacing threshold was 0.2V with a sensed R wave of 9 mv.The distal end of the lead was then tunnelled subcutaneously approximately 6 cm below the clavicle. The lead was sutured to the subcutaneous tissues at the subclavian puncture site with the lead cuff, and this small incision closed with a single suture (Figures 1 and 2). A single-chamber pulse generator (Medtronic Sigma) programmed to bipolar pacing and bipolar sensing was connected to the external end of the pacing lead. Acute implantation parameters were: R wave 10mV, threshold 0.3V at a pulse duration of 0.5 ms. Following implantation of the pacing lead, the device continued to display excellent threshold (0.6V and 0.5V at 1.0ms pulse duration) and sensing values (R wave 10mV, 9mV) at 1 and 3 months respectively, and there was no sign of local infection. The patient responded slowly to parenteral antibiotics, with gradual reduction in the inflammatory markers, and blood cultures became sterile. He subsequently underwent cardiac surgery four months after diagnosis of endocarditis. As the PR interval had shortened in response to medical treatment, and no episodes of high-grade AV block occurred, the pacing lead was ultimately removed two weeks after surgery by freeing the subcutaneous retainer cuff, retracting the fixation helix at the tip of the electrode and by applying gentle traction. The duration of the implant was four months in total.


Figure 2.
Fluoroscopic image in a right anterior oblique view taken at the time of pre-operative cardiac catheterization. The tip of the central venous catheter (Permacath) is in the proximal superior vena cava, and a pigtail catheter used to perform ascending aortography overlies it in this view. The fixation helix can be seen at the end of the pacemaker lead.


Discussion
       This approach to expected long-term temporary pacing proved highly successful in our patient. The procedure was well tolerated and provided long-term pacing support with a lower risk of line infection compared to the use of traditional trans-venous temporary pacing wires and venous sheaths. In addition, the significant risk of complications from temporary pacing is reduced due to avoidance of repeat procedures for repositioning and rotation of venous access sites, particularly in cases such as this where lead stability has been poor. One should always consider this method as an option where temporary pacing is likely to be required for protracted periods.
       It is of interest that the clinical course was complicated by infection of the tunnelled venous cannula, but the pacing system remained free of infection. As the pacemaker generator is not internalized, it is possible to clean and re-use the device in future cases. An important point to emphasize is that the pulse generator must be programmed to bipolar sensing, as the generator is external and the default setting would normally be unipolar sensing from lead-tip to can.
       There are little published data on this technique, although a report on a series of patients described low complication rates.5 This technique could be used more often, and this article will hopefully increase clinicians’ awareness and highlight the relative simplicity of the procedure.


1. Betts TR. Regional survey of temporary transvenous pacing procedures and complications. Postgrad Med J 2003;79:463–465.
2. Austin JL, Preis LK, Crampton RS, et al. Analysis of pacemaker malfunction and complications in the coronary care unit. Am J Cardiol 1982;49:301–306.
3. Winner S, Boon N. Clinical problems with temporary pacemaker prior to permanent pacing. J R Coll Physicians Lond 1989;23:161–163.
4. Hynes JK, Holmes DR Jr, Harrison CE. Five-year experience with temporary pacemaker therapy in the coronary care unit. Mayo Clin Proc 1983;58:122–126.
5. Lever N, Ferguson JD, Bashir Y, Channon KM. Prolonged temporary cardiac pacing using subcutaneous tunnelled active-fixation permanent pacing leads using subcutaneous tunnelled active-fixation permanent pacing leads. Heart 2003;89:209–210.

EP Lab Digest - ISSN: 1535-2226 - Volume 5 - Issue 10 (Oct.2005) - October 2005 - Pages: 22 -

© 2008 HMP Communications
|
All Rights Reserved
83 General Warren Blvd, Suite 100
|
Malvern, PA 19355
Phone: 610-560-0500
|
Fax: 866-488-8273
NACCME.com is your one-stop source for continuing education. Browse through archived webcasts, journal articles, as well as upcoming live events and symposia at www.naccme.com/cardiology














Search Articles




Contact Us

  • Subscribe to EP Lab Digest
  • Article Submission
  • Advertise with Us
  • Become a Spotlight Interview
  • Email Discussion Group
  • Letter to the Editor
  • Subscribe to EP Lab Enews
EP-AF