Troubleshooter's Case Book for Pacemaker and ICD Follow-Up: Fusion, Pseudofusion and Lots of Confusion: It's All About Timing!

A textbook unpaced ECG from a healthy patient. Notice that the QRS complexes are narrow, tall, and sharp. (photo permission courtesy of St. Jude Medical, Inc.)
A ventricular single-chamber pacemaker captures the heart (a depolarization follows immediately after the large, unipolar pacing spikes). Note that the QRS is wider, notched, and more "rounded" than an unpaced complex. (photo permission courtesy
A dual-chamber paced surface ECG as seen on a programmer. The annotations (A and V for paced atrial and paced ventricular events) and timing cycles report what the device is thinking. In this case, the device is pacing at 1,000 ms intervals (which transla
Fusion in a single-chamber paced ECG is indicated in the circled complexes. Notice that the fused complexes do not look like unpaced QRS complexes (which here are small, sharp, and narrow) nor do they look like paced QRS complexes (which are wider with pr
The circled complex represents pseudofusion, where a pacemaker spike falls on top of an intrinsic beat. Unlike fusion (where spike and intrinsic activity both contribute to the depolarization), the spike makes no contribution at all to a pseudofusion beat
This stylized ECG shows a fused atrial beat in the last complex. Like ventricular activity, a fused atrial beat has a unique morphology that is unlike an intrinsic or paced p-wave. (photo permission courtesy of St. Jude Medical, Inc.)
This surface ECG shows seven complexes. The first complex reveals a paced QRS complex with a characteristic wide, notched morphology. The second complex looks suspiciously like fusion, mainly because of the variation in QRS morphology. The third, fourth,
Ready to Take the Fusion and Pseudofusion Challenge?  There are eight complexes on this VVI rhythm strip. Which complexes make you think you might be dealing with fusion, and which make you think that you might be dealing with pseudofusion? What do you d
Author(s): 

Tom Kenny, RN

First, let's review normal conduction in the healthy heart (Figure 1). The heart s natural pacemaker in the sinoatrial (SA) node delivers an electrical output which conducts down over the atria, regroups in the AV node, and then travels over the ventricles. Because the atria are small, their depolarization is captured on the surface ECG as the relatively small p-wave. The ventricles depolarization is a much larger electrical event and is recorded on the surface ECG with tall, sharp QRS complexes.

A pacemaker changes the surface ECG in several predictable ways (Figure 2). The pacemaker output pulse appears on a paced ECG as a vertical line or "spike," which is typically taller in unipolar systems versus bipolar systems. The pacemaker spike should result in an immediate depolarization. But when a ventricular output "captures" or causes a depolarization, the resulting QRS complex is wider and often exhibits a characteristic notched effect, typical of unpaced patients with left bundle-branch block (LBBB).

Most clinicians, including me, prefer to work with surface ECGs simply because we are more familiar with them. But in device-based therapy, we frequently have to deal with surface ECGs from the device programmer (Figure 3), which present us with a more condensed image. But neither of these is what the device "sees"! The device relies on the intracardiac electrogram or IEGM, which comes from electrodes within the heart. Interpreting the IEGM is probably the most accurate way of understanding what the device is "thinking." The next best thing is to rely on the annotations that appear on the programmer surface ECG. These codes and numbers report how the device interprets events.

Fusion and pseudofusion have a confusing appearance on the surface ECG. Fusion occurs when a device output spike appears to collide with an intrinsic event. When this happens, it is probably unclear as to what is actually going on (Figure 4). For example, you should be wondering if there is a sensing problem. After all, if there was an intrinsic event, shouldn t the device have "seen" it and inhibited the pacemaker output? Then again, when you look closely, maybe the output and the intrinsic event occurred practically simultaneously. Is that a problem?

When a pacemaker output occurs almost simultaneously with an intrinsic event, the result is called fusion. In such cases, both the pacemaker and the intrinsic activity contribute to the depolarization. The result is a unique kind of beat with a unique-looking QRS complex. However, fusion is not the most efficient form of device therapy. Most device experts will tell you that for conventional pacing, it is best for intrinsic behavior to prevail whenever possible; fusion prevents this. Furthermore, fusion is wasteful because an unnecessary pacemaker output is delivered.

While fused beats do have a unique look, you should never rely on "eyeballing" a surface ECG for your diagnosis. First of all, the surface ECG is not what the device "sees," and second, it is easy to be deceived. While you should always suspect fusion in such cases, it is smart to go back to the basics.

When you have fusion, the one thing you know you do have is capture. The spike is followed by a depolarization. Can the device sense intrinsic activity properly? An easy way to confirm proper sensing is to temporarily decrease the pacing rate long enough to confirm that when pacemaker spikes do appear, the device can see them and inhibit the next output.

image description image description


Anonymoussays: April 22.2010 at 13:44 pm

Pseudofusion and fusion is common in the field and is often frustrating to the reps to diagnose. The ekg appears to be an intrinsic rhtyhm, yet the device "thinks" it has paced/captured...therefore is happy. But, that is certainly not the case. It is always best to ask yourself "What is the device seeing?" It's not as important for you to see what is going on through the surface ekg, but rather what the intracardiacs are allowing the pacemaker/icd to "see."

This is a great article and I appreciate a forum to read about this. Thank you.

-Pacer Rep in SoCal

Reply to this comment »

Post new comment

  • Web page addresses and e-mail addresses turn into links automatically.
  • Lines and paragraphs break automatically.

More information about formatting options

Image CAPTCHA
Enter the characters shown in the image.